IGM400 FAQ

The hot cathode Bayard-Alpert ionization vacuum gauge operates by ionizing the gas inside the gauge and then measuring the number of ions generated. The ions are then collected giving a measurement of the density or pressure of the gas inside the transducer. The various electrodes used in the transducer design are a collector surrounded by a circular grid with one or two filaments outside the grid.

An electric current is passed through the filament to cause the filament temperature to increase. As the filament temperature is increased, electrons are emitted from the filament surface. The bias voltage between the filament and the grid will accelerate the electrons toward the grid. Most electrons will pass through the grid volume and exit the other side of the grid and then be drawn back into the grid for another traversal through the grid volume. Eventually, most electrons will impact the grid surface generating a current between the filament and the grid which is referred to as the emission current. The electronic controller is designed to maintain a constant, selectable emission current which is independent of pressure. While an electron is traversing inside the grid volume, it may collide with a gas molecule and ionize it which removes an electron. The ionized gas molecule, which now has a positive charge because it is missing an electron, will be attracted to the collector which is at ground potential. A current will be generated by the flow of ions to the collector which is known as the ion or collector current. An electrometer will measure the small ion current generated and a pressure, which is proportional to the ion current, can be calculated.

IGM401

  • 1.00E-09 to 5.00E-02 Torr
  • 1.30E-09 to 6.70E-02 mbar
  • 1.30E-07 to 6.7 Pa

Yes. The IGM400 is not a functional device without the InstruTech B-RAX or the FlexRax controller.

No. The IGM400 is only functional when used with the InstruTech B-RAX or the FlexRax controller.

Emission current is the quantity of electrons emitted from the filament into the space inside the ion gauge transducer. The emission current can be changed to allow optimized operation of the ion gauge transducer at lower or higher vacuum. In general in higher pressure ranges of 5.00E-06 Torr to 5.00E-02 Torr, the 100 uA emission setting is preferred. At lower operating pressure ranges of 1.00E-09 Torr to 1.00E-04 Torr the 4 mA emission current setting should be used.

There are no usable signals from the IGM400. The IGM400 interfaces with the B-RAX or the FlexRax controller using its own device specific serial communications. See B-RAX or the FlexRax data sheets for available output signals.

InstruTech offers two different filament materials.

  1. Yttria Coated Iridium
  2. Tungsten

For general vacuum applications, yttria coated iridium filaments are offered for use with inert gases such as air, Nitrogen, argon, etc. Tungsten filaments are available for use with gases that may not be compatible with yttria. The other tradeoff between the two filaments is that the tungsten filament will not survive accidental venting to atmosphere whereas yttria will. It is very important to make sure the tungsten filament is turned off before bringing the chamber up to atmosphere, especially if air is being used to vent the chamber.

The IGM400 is offered with standard dual filaments.

You can select which filament to turn on using the front panel push buttons of the B-RAX or the FlexRax controller. We recommend periodic switching of the filaments every few months.

If emission current is set to 100 uA, the filament can be turned on when pressure drops below 5.00E-02 Torr.
If emission current is set to 4 mA, the filament can be turned on when pressure drops below 1.00E-03 Torr.

Yes you can turn on the filament manually using the front panel push buttons of your controller.

You can turn on the filament automatically by configuring the B-RAX or the FlexRax controller to use readings from another gauge such as an InstruTech convection gauge.

Degas is used to rid the gauge sensor of adsorbed gas. Degas is achieved by applying Electron Bombardment (EB) to the grid. The sensor’s low pressure performance will normally improve after each degas cycle.

Degas must be applied while the filament is turned on and operating. Ensure vacuum level is at or less than 5.00E-05 Torr before attempting to initiate degas.

You can initiate degas from the B-RAX or FlexRax controller.